Features
- Touch Panel with Color LCD Display
- 9 Different Measurement Network to Simulate the Resistance of Human Body
- 8 Different Types of Leakage current for Medical & General Electrical Equipment
- Various Leakage Current Measuring Mode:DC/AC/AC+DC/AC Peak
- Upper & Lower Limitation for PASS/FAIL Judgment
- Meter Mode with SELV/CONV Function
- 50 Preset Test Conditions Conform to IEC 60990; 30 Sets Memories for Customer Defined
- Various Standard Interfaces:RS232/GPIB/USB Host & Device/EXT I/O
-
-
General
|
Display
|
320 x 240 Color TFT LCD
|
|
Operation
|
Touch panel
|
|
Measurement Device
|
9 kinds
|
DC
|
Ranges
|
50uA/500uA/5mA/25mA
|
|
Resolution
|
0.01uA/0.1uA/1uA/10uA
|
|
Accuracy
|
±1.0% fs/±(0.2% rdg+3dgt)
|
AC/AC+DC
|
Ranges
|
50uA/500uA/5mA/25mA
|
|
Resolution
|
0.01uA/0.1uA/1uA/10uA
|
|
Accuracy
|
±2.0% fs/±(2.0% rdg+6dgt) -- 10Hz<f≦100kHz
|
|
|
±2.0% fs/±(2.0% rdg+10dgt) -- 100kHz<f≦1MHz
|
AC peak
|
Ranges
|
500uA/1mA/10mA/75mA
|
|
Resolution
|
0.1uA/1uA/10uA/100uA
|
|
Accuracy
|
±4.0%fs/±2.5%fs/±(2.0% rdg+6dgt)
-- 20Hz<f≦1kHz
|
|
|
±5.0%fs/±5.0%fs/±(5.0% rdg+10dgt)
-- 1kHz<f≦10kHz
|
Meter Mode
|
AC/DC/AC+DC
|
10.0~300.0V
|
|
AC peak
|
15.0~430.0V
|
|
SELV function
|
Available
|
EUT (V/I Check)
|
Voltage range
|
85V ~ 300V
|
|
Current range
|
0.5A ~ 10A
|
Memory
|
Empty sets
|
30 sets
|
|
Preset sets
|
50 sets (conform to IEC 60990)
|
Interfaces
|
Front panel
|
USB host
|
|
Rear panel
|
RS-232C/GPIB/EXT I/O/USB device
|
Power Source
|
For GLC-9000
|
AC 100V/120V/220V/230V±10%,50/60Hz
|
|
For EUT
|
AC 85V~250V,50/60Hz
|
Dimensions
|
|
330 (W) x 150 (H) x 350 (D) mm
|
Weight
|
|
Approx. 5kg
|
|
Simple and Intuitive System
|
Various Measurement Network
|

|

|
The color TFT touch screen makes operation intuitive and simple and also makes it easier to observe test result.
|
Nine Measurement Network are available for measuring the leakage current of electrical and medical equipment.
|
|
|
|
|