AC WITHSTANDING
|
Output-Voltage Range
|
0.050kV~ 5.000kV ac
|
Output-Voltage Resolution
|
2V/step
|
Output-Voltage Accuracy
|
± (1% of setting +5V) [no load]
|
Maximum Rated Load
|
200 VA (5kV/40mA)
|
Maximum Rated Current
|
40mA (0.5kV < V ≦ 5kV)
10mA (0.05kV ≦ V≦ 0.5kV)
|
Output-Voltage Waveform
|
Sine wave
|
Output-Voltage Frequency
|
50 Hz / 60 Hz selectable
|
Voltage regulation
|
± (1% of reading +5V) [full load → no load]
|
Voltmeter Accuracy
|
± (1% of reading+ 5V)
|
Current Measurement Range
|
0.001mA~40.0mA
|
Current Best Resolution
|
0.001mA / 0.01mA / 0.1mA
|
AC Current Measurement Accuracy
|
±(1.5% of reading +30 counts) when HISET < 1.11mA
±(1.5% of reading +3 counts) when HISET ≧ 1.11mA
|
Window Comparator Method
|
Yes
|
ARC Detect
|
Yes
|
RAMP (Ramp-Up Time)
|
0.1s~999.9s
|
TIMER (Test Time)*
|
OFF, 0.5s~999.9s
|
Sweep Function*
|
NOT Support
|
GND
|
ON/OFF
|
DC WITHSTANDING
|
Output-Voltage Range
|
0.050kV~ 6.000kV dc
|
Output-Voltage Resolution
|
2V/step
|
Output-Voltage Accuracy
|
± (1% of setting +5V) [no load]
|
Maximum Rated Load
|
50W (5kV/10mA)
|
Maximum Rated Current
|
10mA (0.5kV < V ≦ 6kV)
2mA (0.05kV ≦ V≦ 0.5kV)
|
Voltage regulation
|
± (1% of reading +5V) [full load → no load]
|
Voltmeter Accuracy
|
± (1% of reading+ 5V)
|
Current Measurement Range
|
0.001mA~10.0mA
|
Current Best Resolution
|
0.001mA / 0.01mA / 0.1mA
|
DC Current Measurement Accuracy
|
±(1.5% of reading +30 counts) when HISET < 1.11mA
±(1.5% of reading +3 counts) when HISET ≧ 1.11mA
|
Window Comparator Method
|
Yes
|
ARC Detect
|
Yes
|
RAMP (Ramp-Up Time)
|
0.1s~999.9s
|
TIMER (Test Time)*
|
OFF, 0.5s~999.9s
|
Sweep Function*
|
NOT Support
|
GND
|
ON/OFF
|
INSULATION RESISITANCE
|
Output-Voltage
|
50V~1000V dc
|
Output-Voltage Resolution
|
50V/step
|
Output-Voltage Accuracy
|
± (1% of setting +5V) [no load]
|
Resistance Measurement Range
|
1M Ω~ 9500M Ω
|
|
Test voltage
|
Measurable Range
|
Accuracy
|
|
50V≦V≦450V
|
1 ~ 50M Ω
51 ~ 2000M Ω
|
±(5% of reading +1 count)
±(10% of reading +1 counts)
|
|
500V≦V≦1000V
|
1 ~ 500M Ω
501 ~ 9500M Ω
|
±(5% of reading +1 count)
±(10% of reading +1 count)
|
Window Comparator Method
|
Yes
|
Output Impedance
|
600k Ω
|
RAMP (Ramp-Up Time)
|
0.1s ~999.9s
|
TIMER (Test Time)
|
1s~999.9s
|
GND
|
OFF (fixed)
|
GROUND BOND
|
Output-Current Range
|
03.00A~30.00A ac
|
Output-Current Resolution
|
0.01A
|
Output-Current Accuracy
|
3A≦I≦8A : ± (1% of setting +0.2A)
8A<I≦30A : ± (1% of setting +0.05A)
|
Test-Voltage
|
6Vac max. (open circuit)
|
Test-Voltage Frequency
|
50 Hz / 60 Hz selectable
|
Resistance Measurement Range
|
10m Ω~650.0m Ω
|
Resistance Measurement Resolution
|
0.1m Ω
|
Resistance Measurement Accuracy
|
± (1% of reading +2m Ω)
|
Window Comparator Method
|
Yes
|
TIMER (Test Time)
|
0.5s~999.9s
|
Sweep Function*
|
NOT Support
|
Test Method
|
Four Terminal
|
MEMORY
|
Single Step Memory
|
MANU : 100 blocks
|
Automatic Testing Memory
|
AUTO : 100 blocks, menu per auto : 16
|
INTERFACE
|
RS232
|
Standard
|
USB (Device)
|
Standard
|
GPIB
|
Option
|
Remote Terminal (Front)
|
Standard
|
SIGNAL IO
|
Standard
|
General
|
DISPLY
|
240 x 64 Ice Blue Dot matrix LCD
|
POWER SOURCE
|
AC100V/120V/220V/230V ±10%, 50/60Hz
|
POWER CONSUMPTION
|
Max. 500VA
|
DIMENSIONS & WEIGHT
|
330(W) x 148(H) x 452(D) mm (Max.),
Approx.19kg max.
|